-40%

Zahn Cup Viscosity Flow Cup #1 to #5 Dip Type 44ml Stainless Steel For Paint Oil

$ 107.52

Availability: 41 in stock
  • All returns accepted: Returns Accepted
  • Refund will be given as: Money back or replacement (buyer's choice)
  • Type: Viscosity Cup
  • Condition: New
  • Item must be returned within: 30 Days
  • MPN: Viscosity Cup
  • Country/Region of Manufacture: China
  • Brand: VTSYIQI
  • Restocking Fee: No
  • Return shipping will be paid by: Buyer

    Description

    Zahn Cup Viscosity Flow Cup #1 to #5 Dip Type 44ml Stainless Steel For Paint Oil
    This link includes 5 kinds of viscosity cups
    Introduction:
    ♦Our Zahn Cup is according to the American society for materials laboratory,paint and raw material standard ASTM Numbers D4212-93 standard test method of immersion viscosity cup of relevant regulations,combined with the actual production.
    Main Features
    ♦Suitable for Pigment Coating Paint Finishes Oil lacquer Paint Ink Varnishes Syrup Etc
    ♦Cup volume 44ml
    ♦Cup material: Stainless Steel
    Technical Specifications:
    ♦Model:Zahn 1#+2#+3#+4#+5#
    ♦Pore size(mm):Φ1.93mm Φ2.69mm Φ3.86mm Φ4.39mm Φ5.41mm,Cup volume 44ml
    ♦Measurement range(cSt):5-60cSt 20-250cSt 100-800cSt 200-1200cSt 400-1800cSt
    ♦Flow time(sec):20-80sec
    ♦Cup material:Stainless Steel
    Operation method:
    1. Use a suitable solvent to wipe clean and dry the cup before use (Note: Please pay special attention to cleaning the small holes of the cup)
    2.Choose the appropriate cup size so that the outflow time is controlled between 20 -80 seconds
    3.Adjust the temperature of the filtered sample and the outflow cup to (23 degrees plus or minus 0.5 degrees) or another agreed temperature
    4.Insert your fingers into the small ring on the handle, and quickly and steadily lift the cup body vertically from the sample to be tested
    5.When the cup body is broken away from the liquid surface,it is important to keep the cup body vertical when the liquid flows out,and the distance from the liquid surface should not exceed 150mm.
    6.Stop timing when the first breakpo